DicksonOne v1.13.0

01/29/2026

DicksonOne v1.13.0 introduces major enhancements to the Cobalt X and Cobalt XS data loggers, including sensor support, alarm behavior, reporting, and device synchronization. This release also includes a broad set of stability improvements across the platform.

Description Impact Recommendation

Bug Fixes and Improvements

  • Swapping identical sensors on the same port no longer creates gaps in historical charts; data continuity is preserved.
  • Archived or deleted loggers stop sending alarms and appear as unassigned devices with a new registration code.
  • Alarm setup pages use clearer wording, pause and resume dialogs clarify that changes take effect after the next reading, and audible-alarm warnings include a “more information” prompt.
  • Default CO₂ channel names now display correctly on devices by replacing the subscript character with a standard “2”.
  • Timestamps are now recorded with millisecond precision for more accurate event tracking.
  • The report export interface was enhanced, and background jobs no longer retry indefinitely when an invalid phone number is provided.
  • Duplicate escalated alarms are prevented, and escalation logic has been tightened to block repeat notifications.
  • Fixed an issue where unit conversions were applied multiple times in reports.
  • Added missing translations to ensure valid PIN acknowledgments and sensor-failure acknowledgments are properly logged in the audit trail.
  • Prevented calibration certificates from being merged incorrectly.
  • Added missing tooltip translations on the gateway firmware tab.
  • Updated acknowledgment handling so the “Other” reason code on Cobalt X2 devices is recognized correctly.
  • Resolved silent failures and translation issues during report generation.
  • Corrected alarm pop-ups that displayed incorrect recent readings for universal-sensor channels.
  • Restored the “Archived” label for archived data loggers.
  • Ensured paused audible alerts are correctly sent to devices.
  • Updated alert reports to count warning events accurately and removed warning events from CSV exports
  • Fixed issues where open alarms did not clear after conditions returned to normal.
  • Eliminated delays in alarm notifications for previously posted data on Cobalt X and Cobalt XS devices.
  • Fixed issues with the dashboard’s “Channels Out of Range” widget.
  • Resolved recurring 500-error responses related to the Heat Index channel.
  • Fixed a 504 timeout error when accessing the alarms page.
  • Prevented unnecessary subscription-expiration notifications.
  • Resolved server errors when configuring audible alerts using mmHg units.
  • Prevented duplicate alarms from being created when the “Save Alarm” button is clicked multiple times.
  • Fixed an issue where archived RFL data loggers could not be re-added to an account.
  • Corrected unit-of-measure conversion inconsistencies for hardware alarms, including relays and audible alarms.
  • Resolved an issue where a deleted gateway could not be re-added to an account.
No impact No testing needed

Terms of Use & Pricing Update

  • Our Terms of Use have been refreshed to reflect new 2026 subscription pricing plans.
No impact No testing needed

Features

  • Smart-Sensor and calibration certificates
    • The Sensor Certificates page now displays Smart-Sensor certificates, custom calibration-point certificates, and as-found calibration certificates.
  • Universal 4–20 mA sensor support
    • Cobalt X and Cobalt XS loggers now handle universal 4–20 mA sensors correctly, including mapping, display, and data synchronization.
  • Custom channel names on devices
    • Channel names configured in DicksonOne are now sent to Cobalt X and Cobalt XS devices, so the on-device display matches your account.
  • Dry-contact sensor support
    • Added support for Cobalt X devices with dry-contact channels, including normally open and normally closed configurations.
  • Audible alarms for dry-contact channels
    • Cobalt X devices now emit an audible alert when a dry-contact alarm triggers.
  • NO/NC configuration support
    • DicksonOne now recognizes and displays normally open and normally closed settings sent from devices.
Minor impact Check new functionality

THIS DWE FIRMWARE IS A REQUIRED UPDATE - FW version 2.0.1 includes a new digital certificate. This is a critical update to ensure that loggers continue to communicate with DicksonOne.

Additionally, this firmware includes a delay in the sensor configuration looping. This is a critical update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.

Other enhancements include DWE compatibility with Oceasoft Smart-Sensors and Open/Close Sensors.

Description Impact Recommendation

Bugs

  • Resolved a bug where the logger used an old token to communicate and therefore failed to post data to DicksonOne
  • Resolved a bug where calibration dates had discrepancies
  • Resolved a bug where old unsent readings were associated with a new sensor serial number after sensors were swapped
  • Resolved a bug where configuration packets were being sent from DWE logger while on battery power (NO AC power connected)
  • Resolved a bug with RTC Synchronization related to Non-Sequential Reading Timestamps
  • Resolved a bug where SSL_MAX_CONTENT_LEN is too small for Amazon Certs issue resolved
Low impact No testing needed

Enhancements

  • DWE now sends sensor configuration packets before sending readings
  • Communications are now only allowed over HTTPS (no more option to turn SSL off)
  • A flag has been added to the API packets to identify the current sensor plugged into a data logger
  • x509 TLS certs have been refreshed
  • Rollback settings protection was added
  • Added a “pod_port” field to the readings packet in the API
  • Added SSID field in the main configuration file when DWE logger is communicating on Ethernet
  • Added exponential retries to DWE when DicksonOne is unavailable
  • Updated RTC Fallback Logic with every firmware release
  • Removed channel & BSSID-based scanning
  • Removed code from the firmware related to an old external battery pack no longer offered
Minor impact Check new functionality

Sensor Support

  • Introduced compatibility with Oceasoft PT100 Smart-Sensor
  • Introduced compatibility with Oceasoft Temperature and Humidity Smart-Sensor
  • Introduced compatibility with Oceasoft Co2 Triple Smart-Sensor
  • Introduced compatibility with Oceasoft Universal Smart-Sensor
  • Introduced compatibility with Open/Close Sensors (ROC & RTHMOC)
Minor impact Check new functionality

THIS TWE/TWP FIRMWARE IS A REQUIRED UPDATE - FW version 2.8.3 includes a new digital certificate. This is a critical update to ensure that logger continue to communicate with DicksonOne.

Additionally, this firmware includes a delay in the sensor configuration looping. This is a critical update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.

Other enhancements include TWE/TWP compatibility with Oceasoft Smart-Sensors and Open/Close Sensors.

Description Impact Recommendation

Sensor Support

  • Adds support for Open/Close sensors.
  • Adds support for Oceasoft Triple-CO2 and Universal 4-20mA sensors.
  • Fixed an issue where certain replaceable sensors briefly displayed invalid readings when first plugged into Touchscreen loggers.
  • Calibration dates read from sensors are understood to be in local time for Chicago, IL (CST or CDT depending on the date), and are converted to UTC when stored internally.
  • Calibration dates are transmitted to DicksonOne in UTC.
  • Calibration dates are converted from UTC to the logger’s local time when displayed onscreen.
  • Adds a delay in sensor reconnection logic
No impact No testing needed

Networking

  • Prevents network loops when NTP fails
  • Adds support for alternate DNS addresses when using Ethernet connections.
  • Adds support for Telit wifi modules.
  • Fixed an issue where Touchscreen loggers would not request new registration codes from DicksonOne after an initial code had expired.
  • Fixed an issue where Touchscreen loggers would become stuck sending DNS queries for up to 20 minutes during data catch-up periods.
  • Include blank SSID when communicating on Ethernet
Low impact No testing needed

DicksonOne Communication

  • Adds support for updated DicksonOne API interface. Removes TLS certificates from the old DicksonOne API interface.
  • Eliminates insecure HTTP-only communication with DicksonOne. All communication is now encrypted using TLS 1.2.
Minor impact Check new functionality

TWE/TWP Firmware version 2.9.0 introduces Heat-Index support, OceaSoft Differential Pressure Sensor support, OceaSoft CO2 Triple Sensor support, and addresses bugs.

Description Impact Recommendation

Bugs

  • Fixed issue where data was not stored or sent after swapping sensor ports
  • Fixed issue where PT100 OceaSoft Smart sensor readings were not displayed
  • Fixed issue where universal sensor mapping was not applied to the logger
  • Corrected pod model reporting for OceaSoft sensors running FW 2.7.0+
  • Fixed touchscreen reboot issue caused by Telit module losing MAC address
  • Fixed incorrect PT100 temperature readings below freezing
  • Fixed Telit module legacy AT-command mode MAC address issue
  • Fixed layout issue in text view with custom universal sensor units of measure
  • Fixed sensor reconnection loop that created excessive sensor entries
  • Fixed Wi-Fi connection/disconnection causing hard fault
  • Updated current sensor API
  • Improved hardware alarm accuracy by rounding configuration values to the nearest tenth
  • Improved sensor retry logic to match DWE2 updates
  • Verified and standardized universal sensor resolution and digit display
No impact No testing needed

Features and Enhancements

  • Added support for multi-server configurations
  • Added support for OceaSoft Differential Pressure sensor
  • Added ability to check backup battery health
  • Added support for OceaSoft CO2 Triple sensor
  • Added Heat Index virtual channel
Low impact No testing needed

DicksonOne v1.12.6

08/06/2025

DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.

Description Impact Recommendation

Bug Fixes and Improvements

  • Improved dashboard performance by optimizing query behavior when no data is returned.
  • Fixed an issue where alarms were still being triggered for out-of-range readings recorded while the alarm was paused.
  • Resolved a bug causing “Error 500” on the Heat Index channel after sensor replacement, ensuring proper merging of channel types.
No impact No testing needed

DicksonOne v1.12.5

07/16/2025

DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.

Description Impact Recommendation

Bug Fixes and Improvements

  • Improved job handling by preventing automatic retries on non-actionable jobs, reducing unnecessary processing and system load.
No impact No testing needed

DWE2 Firmware version 1.3.0 introduces the Heat-Index support, optimizes power consumption, and enhances support of the OceaSoft Differential Pressure sensor.

Description Impact Recommendation
  • Added Heat-Index virtual channel
  • Optimized Open-Close sensor power consumption on battery
  • Added support for OceaSoft Differential Pressure Sensor in Pascals
Low impact No testing needed

DicksonOne v1.12.4

07/01/2025

DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.

Description Impact Recommendation

Bug Fixes and Improvements

  • Improved load times for Overview / Custom Dashboard page
  • Optimized alarm job processing
  • Improved system testing by introducing a stress test tool
  • Resolved an issue with Internal 500 Errors on the Billing Page
  • Improved Overall French Translation of DicksonOne Platform
  • Added Metadata to Login Page for Search Engine Optimization 
  • Resolved an issue with the Subscribe Button on the Billing Page
  • Implemented a fix for Alarm Events that fail and retry repeatedly, causing inconsistent alarm events
  • Resolved an issue where phone call alarms were in the incorrect language based on the users settings
  • Resolved an issue where sensor records were removed from the Manage Sensor table after a sensor swap
  • Resolved an issue where Relay Sensor was not displayed when setting up notification policies
  • Resolved an issue where the Warning filter on the Alert Overview page did not accurately filter alerts
  • Removed extra character that was present in CSV export files containing negative values
  • Resolved an issue where Relay Alarms were not automatically sent down to the data logger
  • Implemented a fix to trigger Back to Normal alarm notifications upon receiving the first data point back in specified range
  • Removed Channel ID and included Channel Name from Daily Summary Report
No impact No testing needed

Features

  • Heat-Index for DWE, DWE2, RFL, and Touchscreen data loggers (Software Support Only)
    • Added Heat-Index calculations across the platform, including Monitoring Point Pages (excluding Locations), Alerting, and Reports.
  • Introduce Incorrect PIN Code entry logging on Audit Trail for Cobalt X Data Loggers
    • Any invalid PIN Code entries will be recorded on the Audit Trail of DicksonOne
  • Introduce Translations for Reports
    • Users may select from French or Spanish report languages when configuring their reports
Low impact No testing needed

FW version 1.2.0 focused on optimizing battery backup power consumption. Multiple bug fixes have been included in this version as well.

Description Impact Recommendation

Bugs

  • Resolved an issue where DWE2 received an invalid MAC during the initial ethernet connection.
  • Resolved a bug where logger lost serial number and token during operation.
  • Resolved a bug where logger lost MAC address after power cycling.
  • Resolved an issue where incorrect sensor model and serial number were sent after swapping to a single-channel sensor.
  • Resolved an issue where audible alerts were not triggering due to configuration packet value rounding.
  • Optimized battery backup power consumption.
Low impact No testing needed