For information regarding change log and version number methodology, please click
here.
TWE/TWP Firmware version 2.9.0 introduces Heat-Index support, OceaSoft Differential Pressure Sensor support, OceaSoft CO2 Triple Sensor support, and addresses bugs.
Description |
Impact |
Recommendation |
Bugs
- Fixed issue where data was not stored or sent after swapping sensor ports
- Fixed issue where PT100 OceaSoft Smart sensor readings were not displayed
- Fixed issue where universal sensor mapping was not applied to the logger
- Corrected pod model reporting for OceaSoft sensors running FW 2.7.0+
- Fixed touchscreen reboot issue caused by Telit module losing MAC address
- Fixed incorrect PT100 temperature readings below freezing
- Fixed Telit module legacy AT-command mode MAC address issue
- Fixed layout issue in text view with custom universal sensor units of measure
- Fixed sensor reconnection loop that created excessive sensor entries
- Fixed Wi-Fi connection/disconnection causing hard fault
- Updated current sensor API
- Improved hardware alarm accuracy by rounding configuration values to the nearest tenth
- Improved sensor retry logic to match DWE2 updates
- Verified and standardized universal sensor resolution and digit display
|
No impact |
No testing needed |
Features and Enhancements
- Added support for multi-server configurations
- Added support for OceaSoft Differential Pressure sensor
- Added ability to check backup battery health
- Added support for OceaSoft CO2 Triple sensor
- Added Heat Index virtual channel
|
Low impact |
No testing needed |
DicksonOne v1.12.6
08/06/2025
DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.
Description |
Impact |
Recommendation |
Bug Fixes and Improvements
- Improved dashboard performance by optimizing query behavior when no data is returned.
- Fixed an issue where alarms were still being triggered for out-of-range readings recorded while the alarm was paused.
- Resolved a bug causing “Error 500” on the Heat Index channel after sensor replacement, ensuring proper merging of channel types.
|
No impact |
No testing needed |
DicksonOne v1.12.5
07/16/2025
DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.
Description |
Impact |
Recommendation |
Bug Fixes and Improvements
- Improved job handling by preventing automatic retries on non-actionable jobs, reducing unnecessary processing and system load.
|
No impact |
No testing needed |
DWE2 Firmware version 1.3.0 introduces the Heat-Index support, optimizes power consumption, and enhances support of the OceaSoft Differential Pressure sensor.
Description |
Impact |
Recommendation |
- Added Heat-Index virtual channel
- Optimized Open-Close sensor power consumption on battery
- Added support for OceaSoft Differential Pressure Sensor in Pascals
|
Low impact |
No testing needed |
DicksonOne v1.12.4
07/01/2025
DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.
Description |
Impact |
Recommendation |
Bug Fixes and Improvements
- Improved load times for Overview / Custom Dashboard page
- Optimized alarm job processing
- Improved system testing by introducing a stress test tool
- Resolved an issue with Internal 500 Errors on the Billing Page
- Improved Overall French Translation of DicksonOne Platform
- Added Metadata to Login Page for Search Engine Optimization
- Resolved an issue with the Subscribe Button on the Billing Page
- Implemented a fix for Alarm Events that fail and retry repeatedly, causing inconsistent alarm events
- Resolved an issue where phone call alarms were in the incorrect language based on the users settings
- Resolved an issue where sensor records were removed from the Manage Sensor table after a sensor swap
- Resolved an issue where Relay Sensor was not displayed when setting up notification policies
- Resolved an issue where the Warning filter on the Alert Overview page did not accurately filter alerts
- Removed extra character that was present in CSV export files containing negative values
- Resolved an issue where Relay Alarms were not automatically sent down to the data logger
- Implemented a fix to trigger Back to Normal alarm notifications upon receiving the first data point back in specified range
- Removed Channel ID and included Channel Name from Daily Summary Report
|
No impact |
No testing needed |
Features
- Heat-Index for DWE, DWE2, RFL, and Touchscreen data loggers (Software Support Only)
- Added Heat-Index calculations across the platform, including Monitoring Point Pages (excluding Locations), Alerting, and Reports.
- Introduce Incorrect PIN Code entry logging on Audit Trail for Cobalt X Data Loggers
- Any invalid PIN Code entries will be recorded on the Audit Trail of DicksonOne
- Introduce Translations for Reports
- Users may select from French or Spanish report languages when configuring their reports
|
Low impact |
No testing needed |
FW version 1.2.1 focuses on addressing a bug impacting the sensor table and improving reliability.
Description |
Impact |
Recommendation |
Bugs
- Revises sampling logic to improve reliability
- Adds sensor communication debug information to ping packets
|
Low impact |
No testing needed |
FW version 1.2.0 focused on optimizing battery backup power consumption. Multiple bug fixes have been included in this version as well.
Description |
Impact |
Recommendation |
Bugs
- Resolved an issue where DWE2 received an invalid MAC during the initial ethernet connection.
- Resolved a bug where logger lost serial number and token during operation.
- Resolved a bug where logger lost MAC address after power cycling.
- Resolved an issue where incorrect sensor model and serial number were sent after swapping to a single-channel sensor.
- Resolved an issue where audible alerts were not triggering due to configuration packet value rounding.
- Optimized battery backup power consumption.
|
Low impact |
No testing needed |
DicksonOne v1.12.2
01/16/2025
DicksonOne patch releases focus on addressing system bugs/issues and have no impact on validation. The purpose of a patch release is to restore any features back to specification so that users do not experience further issues with the system.
Description |
Impact |
Recommendation |
Updates to our Terms of Use
- Updated Terms of Use to reflect changes to the Monthly Data Logger pricing for 2025. You can read the entire Dickson Terms of Use here.
|
No impact |
No testing needed |
THIS DWE2 FIRMWARE IS A REQUIRED UPDATE - This firmware includes one bug fix which addresses the sensor configuration looping. This is a critical and important update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
Description |
Impact |
Recommendation |
Bugs
- Implemented a 10-minute delay between disconnecting a sensor and sending an empty
sensor packet to DicksonOne.
|
Low impact |
No testing needed |
THIS RFL FIRMWARE IS A REQUIRED UPDATE IF RFL IS USING FIRMWARE VERSION 1.4.0 OR NEWER - This firmware includes one bug fix which addresses the sensor configuration looping. This is a critical and important update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
Description |
Impact |
Recommendation |
Bugs
- Implemented a 10-minute delay between disconnecting a sensor and sending an empty
sensor packet to DicksonOne.
|
Low impact |
No testing needed |