TWE/TWP Firmware version 2.10.0 improves compatibility with the latest Telit firmware, increases system stability during server outages by reducing unnecessary communication overhead, and resolves several reliability issues.

Description Impact Recommendation

Bugs

  • Resolved an issue where repeated pod swaps could cause device instability over time.
  • Resolved an issue that could cause devices to stop functioning correctly after a limited number of pod swaps.
  • Improved device behavior during server outages by reducing unnecessary communication attempts, lowering the risk of data loss or connection errors.
  • Improved clock accuracy and reliability under fallback conditions.
No impact No testing needed

Features and Enhancements

  • Added support for the latest Telit WiFi module firmware (v39.03.012), ensuring continued compatibility with Touchscreen devices.
No impact No testing needed

THIS TWE/TWP FIRMWARE IS A REQUIRED UPDATE - FW version 2.8.3 includes a new digital certificate. This is a critical update to ensure that logger continue to communicate with DicksonOne.

Additionally, this firmware includes a delay in the sensor configuration looping. This is a critical update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.

Other enhancements include TWE/TWP compatibility with Oceasoft Smart-Sensors and Open/Close Sensors.

Description Impact Recommendation

Sensor Support

  • Adds support for Open/Close sensors.
  • Adds support for Oceasoft Triple-CO2 and Universal 4-20mA sensors.
  • Fixed an issue where certain replaceable sensors briefly displayed invalid readings when first plugged into Touchscreen loggers.
  • Calibration dates read from sensors are understood to be in local time for Chicago, IL (CST or CDT depending on the date), and are converted to UTC when stored internally.
  • Calibration dates are transmitted to DicksonOne in UTC.
  • Calibration dates are converted from UTC to the logger’s local time when displayed onscreen.
  • Adds a delay in sensor reconnection logic
No impact No testing needed

Networking

  • Prevents network loops when NTP fails
  • Adds support for alternate DNS addresses when using Ethernet connections.
  • Adds support for Telit wifi modules.
  • Fixed an issue where Touchscreen loggers would not request new registration codes from DicksonOne after an initial code had expired.
  • Fixed an issue where Touchscreen loggers would become stuck sending DNS queries for up to 20 minutes during data catch-up periods.
  • Include blank SSID when communicating on Ethernet
Low impact No testing needed

DicksonOne Communication

  • Adds support for updated DicksonOne API interface. Removes TLS certificates from the old DicksonOne API interface.
  • Eliminates insecure HTTP-only communication with DicksonOne. All communication is now encrypted using TLS 1.2.
Minor impact Check new functionality

TWE/TWP Firmware version 2.9.0 introduces Heat-Index support, OceaSoft Differential Pressure Sensor support, OceaSoft CO2 Triple Sensor support, and addresses bugs.

Description Impact Recommendation

Bugs

  • Fixed issue where data was not stored or sent after swapping sensor ports
  • Fixed issue where PT100 OceaSoft Smart sensor readings were not displayed
  • Fixed issue where universal sensor mapping was not applied to the logger
  • Corrected pod model reporting for OceaSoft sensors running FW 2.7.0+
  • Fixed touchscreen reboot issue caused by Telit module losing MAC address
  • Fixed incorrect PT100 temperature readings below freezing
  • Fixed Telit module legacy AT-command mode MAC address issue
  • Fixed layout issue in text view with custom universal sensor units of measure
  • Fixed sensor reconnection loop that created excessive sensor entries
  • Fixed Wi-Fi connection/disconnection causing hard fault
  • Updated current sensor API
  • Improved hardware alarm accuracy by rounding configuration values to the nearest tenth
  • Improved sensor retry logic to match DWE2 updates
  • Verified and standardized universal sensor resolution and digit display
No impact No testing needed

Features and Enhancements

  • Added support for multi-server configurations
  • Added support for OceaSoft Differential Pressure sensor
  • Added ability to check backup battery health
  • Added support for OceaSoft CO2 Triple sensor
  • Added Heat Index virtual channel
Low impact No testing needed

This release solely adds support for a new component on the circuit boards of the touchscreen models. This firmware is backwards compatible with old revisions of the circuit board but offer no user-facing functional enhancements.

Description Impact Recommendation

Support for Component Replacements

  • Added backwards compatible support for new flash memory chip
  • Updated bootloader to support new flash chip
No impact No testing needed
Description Impact Recommendation

Enhancements

  • Updated SSL Cipher Suite
  • Implement Additional Differential Pressure Units
  • Audible Calibration Alarm was replaced with a calibration alarm pop up
  • Downloading data from TSB using DicksonWare 2.0 with USB cable
  • Add Sensor Port to Readings Packet
  • Added backup settings to prevent devices from resetting their own settings
Low impact No testing needed

Bugs

  • Enterprise Connection attempts Authentication with Username GSN
  • Alarms triggered at alarm set value
  • Sensor swaps caused device reboot
  • Logger would not report model SKU to DicksonOne
  • Calibration warnings on the device displayed the wrong port number
  • Clearing the Min/Max with a probe error caused high Min/Max values
  • Device does not boot up from a low power state
  • Parse Incoming Proxy Settings through API
  • New channels are not created properly when swapping different sensor models
  • Device does not send readings packets while still sending ping and configuration packets
  • Units lock up and reboot
Minor impact Check new functionality
Description Impact Recommendation
  • BSSID to be included in API logs
  • Option to export Partial (Since Last Export) or all Historical Data (Entire History) have been added
  • USB icon flashes on the display when taking a screenshot
  • Min/Max has been added to the Text Display View
  • A button to reset the Min and Max has been added
  • Viewable characters have been expanded for the device name on the screen
  • Dallas/Sensiron Pod Replacement to improve sensor swapping
  • Connection improvement has made to adhere to Access Point hopping based on RSSI values
  • WiFi Scan List has been improved to display more SSIDs
  • Thread IDs are now reset properly when performing a factory reset on the device
  • The label for Device Code has been renamed to Registration Code to mirror DicksonOne
No impact No testing needed
Description Impact Recommendation
  • Adjust real-time clock’s offset to account for drift (mainly for TSBs or TWE/TWPs used without DicksonOne)
  • Changed default sample rate to 5 minutes instead of 1 minute
No impact No testing needed
  • Resolved bug where battery may be drained too quickly.
  • Resolved bug where USB exports attributed readings to the incorrect serial number
Low impact No testing needed

This update is required for continued operation of your TWE/TWP after October 12, 2018.

Description Impact Recommendation

This is the same as the pervious version (2.4.27) but has the new security certificates to address Google’s decision to distrust some existing certificates.

No impact No testing needed
Description Impact Recommendation
  • Patch for graph plotting and USB data export
  • Update to Ping/Config Packet interval during catch-up
  • Patch for Proxy over Ethernet with HTTPS functionality
  • General improvements to Ethernet communication (non-proxy)
  • Add ping packet button popups
  • Implement differential pressure functionality
  • Update for displaying differential pressure data on display and USB export
  • Adjustment to the humidity min/max custom setting button
  • Update to differential pressure .csv export to display data up to 3 decimal places
  • Update to diff pressure .loj export to enable data up to 3 decimal places
  • Data export to Flash drive (Differential Pressure)
  • Implement validation to verify a block is empty and available for immediate use
  • Enable backwards compatibility with 2.1.4 (Current Production Version)
No impact No testing needed
Description Impact Recommendation
  • Fixed issue causing extended decimal places in readings.

  • Changed network layer to prioritize sending readings.

  • Added additional logging to ping packet.

  • Increased watchdog timeout for network layer.

  • BSSID and SSID now included in debug information.

  • SSID and DHCP now included in network settings packet.

  • Detection of board revision and updated power control logic.

  • Reduced packet frequencies of ping and config. Added button on network settings page to send a ping packet.

  • Removed block on having multiple multi-channel pods attached.

  • Fix for connection to WEP networks.

  • Fixed bug in data export of dicksonware file.

No impact No testing needed