For information regarding change log and version number methodology, please click
here.
TWE/TWP Firmware version 2.10.0 improves compatibility with the latest Telit firmware, increases system stability during server outages by reducing unnecessary communication overhead, and resolves several reliability issues.
| Description |
Impact |
Recommendation |
Bugs
- Resolved an issue where repeated pod swaps could cause device instability over time.
- Resolved an issue that could cause devices to stop functioning correctly after a limited number of pod swaps.
- Improved device behavior during server outages by reducing unnecessary communication attempts, lowering the risk of data loss or connection errors.
- Improved clock accuracy and reliability under fallback conditions.
|
No impact |
No testing needed |
Features and Enhancements
- Added support for the latest Telit WiFi module firmware (v39.03.012), ensuring continued compatibility with Touchscreen devices.
|
No impact |
No testing needed |
THIS DWE FIRMWARE IS A REQUIRED UPDATE - FW version 2.0.1 includes a new digital certificate. This is a critical update to ensure that loggers continue to communicate with DicksonOne.
Additionally, this firmware includes a delay in the sensor configuration looping. This is a critical update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
Other enhancements include DWE compatibility with Oceasoft Smart-Sensors and Open/Close Sensors.
| Description |
Impact |
Recommendation |
Bugs
- Resolved a bug where the logger used an old token to communicate and therefore failed to post data to DicksonOne
- Resolved a bug where calibration dates had discrepancies
- Resolved a bug where old unsent readings were associated with a new sensor serial number after sensors were swapped
- Resolved a bug where configuration packets were being sent from DWE logger while on battery power (NO AC power connected)
- Resolved a bug with RTC Synchronization related to Non-Sequential Reading Timestamps
- Resolved a bug where SSL_MAX_CONTENT_LEN is too small for Amazon Certs issue resolved
|
Low impact |
No testing needed |
Enhancements
- DWE now sends sensor configuration packets before sending readings
- Communications are now only allowed over HTTPS (no more option to turn SSL off)
- A flag has been added to the API packets to identify the current sensor plugged into a data logger
- x509 TLS certs have been refreshed
- Rollback settings protection was added
- Added a “pod_port” field to the readings packet in the API
- Added SSID field in the main configuration file when DWE logger is communicating on Ethernet
- Added exponential retries to DWE when DicksonOne is unavailable
- Updated RTC Fallback Logic with every firmware release
- Removed channel & BSSID-based scanning
- Removed code from the firmware related to an old external battery pack no longer offered
|
Minor impact |
Check new functionality |
Sensor Support
- Introduced compatibility with Oceasoft PT100 Smart-Sensor
- Introduced compatibility with Oceasoft Temperature and Humidity Smart-Sensor
- Introduced compatibility with Oceasoft Co2 Triple Smart-Sensor
- Introduced compatibility with Oceasoft Universal Smart-Sensor
- Introduced compatibility with Open/Close Sensors (ROC & RTHMOC)
|
Minor impact |
Check new functionality |
THIS TWE/TWP FIRMWARE IS A REQUIRED UPDATE - FW version 2.8.3 includes a new digital certificate. This is a critical update to ensure that logger continue to communicate with DicksonOne.
Additionally, this firmware includes a delay in the sensor configuration looping. This is a critical update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
Other enhancements include TWE/TWP compatibility with Oceasoft Smart-Sensors and Open/Close Sensors.
| Description |
Impact |
Recommendation |
Sensor Support
- Adds support for Open/Close sensors.
- Adds support for Oceasoft Triple-CO2 and Universal 4-20mA sensors.
- Fixed an issue where certain replaceable sensors briefly displayed invalid readings when first plugged into Touchscreen loggers.
- Calibration dates read from sensors are understood to be in local time for Chicago, IL (CST or CDT depending on the date), and are converted to UTC when stored internally.
- Calibration dates are transmitted to DicksonOne in UTC.
- Calibration dates are converted from UTC to the logger’s local time when displayed onscreen.
- Adds a delay in sensor reconnection logic
|
No impact |
No testing needed |
Networking
- Prevents network loops when NTP fails
- Adds support for alternate DNS addresses when using Ethernet connections.
- Adds support for Telit wifi modules.
- Fixed an issue where Touchscreen loggers would not request new registration codes from DicksonOne after an initial code had expired.
- Fixed an issue where Touchscreen loggers would become stuck sending DNS queries for up to 20 minutes during data catch-up periods.
- Include blank SSID when communicating on Ethernet
|
Low impact |
No testing needed |
DicksonOne Communication
- Adds support for updated DicksonOne API interface. Removes TLS certificates from the old DicksonOne API interface.
- Eliminates insecure HTTP-only communication with DicksonOne. All communication is now encrypted using TLS 1.2.
|
Minor impact |
Check new functionality |
TWE/TWP Firmware version 2.9.0 introduces Heat-Index support, OceaSoft Differential Pressure Sensor support, OceaSoft CO2 Triple Sensor support, and addresses bugs.
| Description |
Impact |
Recommendation |
Bugs
- Fixed issue where data was not stored or sent after swapping sensor ports
- Fixed issue where PT100 OceaSoft Smart sensor readings were not displayed
- Fixed issue where universal sensor mapping was not applied to the logger
- Corrected pod model reporting for OceaSoft sensors running FW 2.7.0+
- Fixed touchscreen reboot issue caused by Telit module losing MAC address
- Fixed incorrect PT100 temperature readings below freezing
- Fixed Telit module legacy AT-command mode MAC address issue
- Fixed layout issue in text view with custom universal sensor units of measure
- Fixed sensor reconnection loop that created excessive sensor entries
- Fixed Wi-Fi connection/disconnection causing hard fault
- Updated current sensor API
- Improved hardware alarm accuracy by rounding configuration values to the nearest tenth
- Improved sensor retry logic to match DWE2 updates
- Verified and standardized universal sensor resolution and digit display
|
No impact |
No testing needed |
Features and Enhancements
- Added support for multi-server configurations
- Added support for OceaSoft Differential Pressure sensor
- Added ability to check backup battery health
- Added support for OceaSoft CO2 Triple sensor
- Added Heat Index virtual channel
|
Low impact |
No testing needed |
DWE2 Firmware version 1.3.0 introduces the Heat-Index support, optimizes power consumption, and enhances support of the OceaSoft Differential Pressure sensor.
| Description |
Impact |
Recommendation |
- Added Heat-Index virtual channel
- Optimized Open-Close sensor power consumption on battery
- Added support for OceaSoft Differential Pressure Sensor in Pascals
|
Low impact |
No testing needed |
FW version 1.2.1 focuses on addressing a bug impacting the sensor table and improving reliability.
| Description |
Impact |
Recommendation |
Bugs
- Revises sampling logic to improve reliability
- Adds sensor communication debug information to ping packets
|
Low impact |
No testing needed |
FW version 1.2.0 focused on optimizing battery backup power consumption. Multiple bug fixes have been included in this version as well.
| Description |
Impact |
Recommendation |
Bugs
- Resolved an issue where DWE2 received an invalid MAC during the initial ethernet connection.
- Resolved a bug where logger lost serial number and token during operation.
- Resolved a bug where logger lost MAC address after power cycling.
- Resolved an issue where incorrect sensor model and serial number were sent after swapping to a single-channel sensor.
- Resolved an issue where audible alerts were not triggering due to configuration packet value rounding.
- Optimized battery backup power consumption.
|
Low impact |
No testing needed |
THIS DWE2 FIRMWARE IS A REQUIRED UPDATE - This firmware includes one bug fix which addresses the sensor configuration looping. This is a critical and important update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
| Description |
Impact |
Recommendation |
Bugs
- Implemented a 10-minute delay between disconnecting a sensor and sending an empty
sensor packet to DicksonOne.
|
Low impact |
No testing needed |
THIS RFL FIRMWARE IS A REQUIRED UPDATE IF RFL IS USING FIRMWARE VERSION 1.4.0 OR NEWER - This firmware includes one bug fix which addresses the sensor configuration looping. This is a critical and important update to ensure that sensor tables do not contain incorrect disconnect and reconnect entries. Please update immediately.
| Description |
Impact |
Recommendation |
Bugs
- Implemented a 10-minute delay between disconnecting a sensor and sending an empty
sensor packet to DicksonOne.
|
Low impact |
No testing needed |
| Description |
Impact |
Recommendation |
This release marks the initial product release for the RFG-US-003 LoRa Gateway. Functionally, the behavior is identical to the RFG-US-002 but changes to the hardware required a new model number and therefore firmware.
|
No impact |
No testing needed |